Author Type

Graduate Student

Date of Award

2012

Document Type

Thesis

Publication Status

Version of Record

Submission Date

October 2026

Abstract

The Thesis explores additional applications of LAP's Aquarius external laser alignment verification Phantom by examining geometric accuracy of magnetic resonance images commonly used for planning intracranial stereotactic radiation surgery (ICSRS) cases. The scans were performed with MRI protocols used for ICSRS, and head and neck diagnosis, and their images fused to computerized tomographic (CT) images. The geometric distortions (GDs) were measured against the CT in all axial, sagittal, and coronal directions at different levels. Using the Aquarius Phantom, one is able to detect GD in ICSRS planning MRI acquisitions, and align the external LAP patient alignment lasers, by following the LAP QA protocol. GDs up to about 2 mm are observed at the distal regions of the longitudinal axis in the SRS treatment planning MR images. Based on the results, one may recommend the use of the Aquarius Phantom to determine if margins should be included for SRS treatment planning.

Extent

xvii, 73 p. : ill. (some col.)

Comments

by Daniel Vergara.|Thesis (M.S.)--Florida Atlantic University, 2012.|Includes bibliography.|Electronic reproduction. Boca Raton, Fla., 2012. Mode of access: World Wide Web.|by Daniel Vergara.

Identifier

FADT3352878

Format

pdf

Share

COinS