Abstract

MapCheck measurements for 50 retrospective patient’s treatment plans suggested that MapCheck could be effectively employed in routine patient specific quality assurance in M6 Cyberknife with beams delivered at different treatment angles. However, these measurements also suggested that for highly intensity modulated MLC plans, field segments of width < 8 mm should further be analyzed with a modified (-4%) correction factor. Results of MC simulations of the M6 Cyberknife using the EGSnrc program for 2-5 millions of incident particles in BEAMnrc and 10-20 millions in DOSXYZnrc have shown dose uncertainties within 2% for open fields from 7.6 x 7.7 mm2 to 100 x 100 mm2. Energy and corresponding FWHM were optimized by comparing with water phantom measurements at 800 mm SAD resulting to E = 7 MeV and FWHM = 2.2 mm. Good agreement of dose profiles (within 2%) and outputs (within 3%) were found between the MC simulations and water phantom measurements for the open fields.