Abstract

Since the release of the Cyberknife Multileaf Collimator (CK-MLC), it has been a constant concern on the realistic dose differences computed with its early-available Finite Size Pencil Beam algorithm (FSPB) from those computed by using industry well-accepted algorithms such as the Monte Carlo (MC) dose algorithm. In this study dose disparities between FSPB and MC dose calculation algorithms for selected CK-MLC treatment plans were quantified. The dosimetry for planning target volume (PTV) and major organs at risks (OAR) was compared by calculating normalized percentage deviations (Ndev) between the two algorithms. It is found that the FSPB algorithm overestimates D95 of PTV when compared with the MC algorithm by averaging 24.0% in detached lung cases, and 15.0% in non-detached lung cases which is attributed to the absence of heterogeneity correction in the FSPB algorithm. Average dose differences are 0.3% in intracranial and 0.9% in pancreas cases. Ndev for the D95 of PTV range from 8.8% to 14.1% for the CK-MLC lung treatment plans with small field (SF ≤ 2x2cm2). Ndev is ranged from 0.5-7.0% for OARs.