Abstract
The purpose of this study is to verify and validate the dose at various points of interest in accelerated partial breast irradiation (APBI) treated with the Strut Adjusted Volume Implant (SAVI) applicator using Thermoluminescent Dosimeters (TLDs). A set of CT images were selected from a patient’s data who had received APBI using the SAVI applicator. The images were used to make 3D models. TLDs were calibrated for Brachytherapy. Various points of interest were marked out and slots were carved in the 3D models to fit the TLDs. CT scans were taken of the 3D models with expanded SAVI applicator inserted. A plan was made following B-39 protocol. The TLDs were read and the absorbed doses were calculated and compared to the delivered doses. The results of this study show that the overall average reading of the TLDs is within expected value. The TPS shows overestimated dose calculations for brachytherapy.