Author

Author Type

Graduate Student

Date of Award

2014

Document Type

Thesis

Publication Status

Version of Record

Submission Date

October 2026

Thesis/Dissertation Advisor [Chair]

Pella, Silvia

Abstract

Common methods for commissioning linear accelerators often neglect beam data for small fields. Examining the methods of beam data collection and modeling for commissioning linear accelerators revealed little to no discussion of the protocols for fields smaller than 4 cm x 4 cm. This leads to decreased confidence levels in the dose calculations and associated monitor units (MUs) for Intensity Modulated Radiation Therapy (IMRT). The parameters of commissioning the Novalis linear accelerator (linac) on the Eclipse Treatment Planning System (TPS) led to the study of challenges collecting data for very small fields. The focus of this thesis is the examination of the protocols for output factor collection and their impact on dose calculations by the TPS for IMRT treatment plans. Improving output factor collection methods, led to significant improvement in absolute dose calculations which correlated with the complexity of the plans.

Extent

109 p.

Comments

Includes bibliography.|Thesis (M.S.)--Florida Atlantic University, 2014.|FAU Electronic Theses and Dissertations Collection|Includes bibliography.