A high sensitivity capacitance dilatometer constructed solely of fused quartz was designed to study phase transitions of sub-millimeter sized samples from 4 to 300 K. The minimal thermal expansion contribution from quartz reduced the overall magnitude of the background by a factor of 10. As a consequence, the increased resolution of the data provided greater sensitivity than has been capable with metallic capacitance dilatometers. A one millimeter long single crystal of La1.67Sr0.33NiO 4 was analyzed from 80 to 300 K. Our highly reproducible measurements reveal phase transitions at similar temperatures to those observed with magnetization and specific heat measurement techniques.
Extent
90 p.
Recommended Citation
Timmins, George Emerson, "Design and construction of a high-sensitivity capacitance-based quartz dilatometer" (2002). Electronic Theses and Dissertations 2000 - 2009. 367.