Abstract
Evaluation of dose optimization using the Pencil Beam (PB) and Monte Carlo (MC) algorithms may allow physicists to apply dosimetric offsets to account for inaccuracies of the PB algorithm for lung cancer treatment with Stereotactic Body Radiotherapy (SBRT). 20 cases of Non-Small Cell Lung Cancer (NSCLC) were selected. Treatment plans were created with Brainlab iPlanDose® 4.1.2. The D97 of the Planning Target Volume (PTV) was normalized to 50 Gy on the Average Intensity Projection (AIP) using the fast PB and compared with MC. This exact plan with the same beam Monitor Units (MUs) was recalculated over each respiratory phase. The results show that the PB algorithm has a 2.3-2.4% less overestimation at the maximum exhalation phase than the maximum inhalation phase when compared to MC. Significantly smaller dose difference between PB and MC is also shown in plans for peripheral lesions (7.7 ± 0.7%) versus central lesions (12.7±0.8%)(p< 0.01).
Extent
83 p.
Recommended Citation
Evans, Grant, "Exploring appropriate offset values for pencil beam and Monte Carlo dose optimization in lung stereotactic body radiotherapy encompassing the effects of respiration and tumor location" (2014). Electronic Theses and Dissertations 2010 - 2019. 907.