Abstract
Intracavitary high dose rate (HDR) brachytherapy is a form of radiation therapy generally in which a post-surgical tissue margin is treated. The dose gradient of HDR brachytherapy is very steep, and thus small displacements of the applicator, even as small as 1 mm, could potentially cause significant variations of dose which could result in undesired side effects such as overdose of a critical organ. In this retrospective dosimetric study, the variation of dose due to various small range motions of gynecological applicators is investigated. The results show that the implementation of additional immobilization and localization devices along with other safety measures needs to be further investigated.
Extent
79 p.
Recommended Citation
Hyvärinen, Mikko, "The importance of immobilization and localization of gynecological applicators in high dose rate brachytherapy treatments" (2015). Electronic Theses and Dissertations 2010 - 2019. 1340.