Author Type

Graduate Student

Date of Award

2015

Document Type

Thesis

Publication Status

Version of Record

Submission Date

October 2026

Thesis/Dissertation Advisor [Chair]

Ouhib, Zoubir

Abstract

Monte Carlo (MC) and Pencil Beam (PB) calculations are compared to their measured planar dose distributions using a 2-D diode array for lung Stereotactic Body Radiation Therapy (SBRT). The planar dose distributions were studied for two different phantom types: an in-house heterogeneous phantom and a homogeneous phantom. The motivation is to mimic the human anatomy during a lung SBRT treatment and incorporate heterogeneities into the pre-treatment Quality Assurance process, where measured and calculated planar dose distributions are compared before the radiation treatment. Individual and combined field dosimetry has been performed for both fixed gantry angle (anterior to posterior) and planned gantry angle delivery. A gamma analysis has been performed for all beam arrangements. The measurements were obtained using the 2-D diode array MapCHECK 2™.

Extent

139 p.

Comments

Includes bibliography.|Thesis (M.S.)--Florida Atlantic University, 2015.|FAU Electronic Theses and Dissertations Collection|Includes bibliography.

Identifier

FA00004360

Format

pdf

Share

COinS