Abstract
An algorithm to determine IMRT optimization parameters within the Elekta Monaco® treatment planning system that increases dose homogeneity and dose conformity in the planning target volume was developed. This algorithm determines IMRT optimization parameters by calculating the difference between two pairs of dose points along the target volume’s dose volume histogram: Dmax – Dmin, and D2 – D98. The algorithm was tested on the Elekta Monaco® Treatment Planning System at GenesisCare of Coconut Creek, Florida using CT data from 10 anonymized patients with non-small cell lung cancer of various tumor sizes and locations. Nine iterations of parameters were tested on each patient. Once the ideal parameters were found, the results were evaluated using the ICRU report 83 homogeneity index as well as the Paddick conformity index. As an outcome of this research, it is recommended that at least three iterations of IMRT optimization parameters should be calculated to find the ideal parameters.
Extent
125 p.
Recommended Citation
Gregorisch, David, "An Algorithm to Determine IMRT Optimization Parameters within the Elekta Monaco® Treatment Planning System that Increases Dose Homogeneity and Dose Conformity in the Planning Target Volume" (2022). Electronic Theses and Dissertations 2020 - Present. 870.
https://digitalcommons.fau.edu/etd_general/870
Identifier
FA00013990
Comments
Includes bibliography.|Thesis (PMS)--Florida Atlantic University, 2022.|FAU Electronic Theses and Dissertations Collection|Includes bibliography.