Author

Author Type

Graduate Student

Date of Award

2021

Document Type

Thesis

Publication Status

Version of Record

Submission Date

October 2026

Thesis/Dissertation Advisor [Chair]

Shang, Charles

Thesis/Dissertation Co-Chair

Muhammad, Wazir

Abstract

In proton therapy systems with pencil-beam scanning, output of Halo effect is not necessarily included in Treatment Planning System (TPS). Halo effect (low-intensity tail) can significantly affect a patient’s dose distribution. The output of this dose depends on the field size being irradiated. Although much research has been made to investigate such relation to the field size, the number of reports on dose calculations including the halo effect is small. In this work we have investigated the Halo effect, including field size factor, target depth factor, and air gaps with a range shifter for a Varian ProBeam. Dose calculations created on the Eclipse Treatment Planning System (vs15.6 TPS) are compared with plane-parallel ionization chambers (PTW Octavius 1500) measurements using PCS and AcurosPT MC model in different isocenters: 5cm, 10cm, and 20cm. We find that in AcurosPT algorithm deviations range between -7.53% (for 2cm field in 25cm air gap with range shifter) up to +7.40% (for 20cm field in 15cm air gap with range shifter). Whereas, in PCS algorithm the deviations are -2.07% (for 20x20cm field in open conditions) to -6.29% (for 20x20cm field in 25cm air gap with range shifter).

Extent

82 p.

Comments

Includes bibliography.|Thesis (M.S.)--Florida Atlantic University, 2021.|FAU Electronic Theses and Dissertations Collection|Includes bibliography.

Identifier

FA00013788